Using
and Operating
Fit trinocular
tube part on bending arm, tighten the screw.
Screw off
protecting cover; install various magnification objectives on
rotating nosepiece in alphabetical, insert – needed eyepiece
in eyepiece tube.
Lay specimen
flake on double – deck stage, clamp down by specimen holding
device, and move to center of view with moving knob.
Close –
up source (AC 220V or AC 110V) turn on switch, regulate handle,
and can be obtained needing brightness of observation. According
to needing to choose appropriate filter.
Adjustment
of pupil distance
Move regulating slide of pupil distance, adjust interval of
eyepiece tube, in order a suitable pupil distance of observer.
This distance can be read among scale of regulating slide.
Turn pupil distance regulating ring of eyepiece tube, conform
the reading number to pupil distance.
Turn iris
diaphragm, continuously change aperture angle of condenser,
fit it on numerical aperture of objective, condenser can rise
and descend. Optical axis of condenser must coincide with duplication
optical axis of objective (has been rectified). If it had deviation,
could adjust three screws on condenser supporter to revise.
First, use
10x objective, to observe, turn coarse focusing knob, up to
appear confusing image at field of view, then turn fine focusing
knob, make the image clear and precise, after that adjust other
high-power objective (100x objective needs use cedar oil). This
time, only need to adjust fine focusing knob can obtain clear
image. 10x, 100x objectives have spring laded, can prevent it
colliding with section.
Adjust the
limit.
Get appropriate degree of tightness and Loss ness of the coarse
focusing knob when turn holding adjustment knob. Note: don’t
adjust too tight, in order to avoid the stage dropping or locked.
Stopper knob can provide fixed clear position of observation,
as well can avoid objective touching specimen flake. The stage
will be reached a fixed position when turn upon stopper knob.
By this time, the position is upper limited of coarse knob,
at this condition, can use fine focusing knob to adjust. After
use, stage ought to descend to its original position.
Pull out
handle, could attach camera or photograph sets on trinocular
tube part.
Change lamp
and fuse.
To screw
on lamp support under electric appliance case, pull out lamp
support could change lamp, screw off fuse cover, can change
fuse.
Maintenance